Below is a Glossary list of terms that may be used in discussions of ethylene oxide (EO) sterilization as recognized by medical device manufacturers inspected by the U.S. FDA. All definitions or explanations are to be taken as applied specifically in the context of EO sterilization processing.
Aeration
Part of the sterilization process during which ethylene oxide and/or its reaction products desorb (outgas) from the medical device until predetermined levels are reached. This may be performed within the sterilizer and/or in a separate chamber or room.
Aeration Area
Either a chamber or a room in which aeration occurs. The temperature in the aeration area is elevated and controlled to assure that the aeration process is accelerated and repeatable. Air is recirculated in the room to maintain good heat distribution with a portion of the air stripped off and cleansed with the pollution control equipment. The stripping of the air reduces the amount of free ethylene oxide resident in the rooms, thus prevents recontamination of product load.
AAMI
Abbreviation for the Association for the Advancement of Medical Instrumentation. AAMI generates guidelines which are used in the industry for validating, monitoring, and performing routine EO sterilization processes.
ANSI/AAMI/ISO 11135:2007
International Organization for Standardization (ISO) Guidance document which has been adopted and published by AAMI (and ANSI) to address validation and routine control of ethylene oxide sterilization.
Bacteriostasis/Fungistasis Test
Test performed to evaluate the presence of microbial growth inhibiting properties of a health care product. This test assures that any reaction of the growth medium with the different materials used in the manufacture of the device will not mask or prevent growth from any viable organism remaining on the device after being subjected to a sterilization process.
Batch
Defined quantity of bulk, intermediate, or finished product that is intended or purported to be uniform in character and quality, and which has been produced during a defined cycle of manufacture.
Bioburden
Population of viable microorganisms on a raw material, component, finished product, and/or package. Unlike the gamma irradiation process, bioburden data is not utilized in ethylene oxide sterilization to determine the sterilization process. It is collected to quantify the amount of product contamination, which is then compared to the population of the biological indicator used in the EO sterilization process.
Biological Indicator
Inoculated carrier contained within its primary pack ready for use and providing a defined resistance to the specified sterilization process. The biological indicator referenced in the United States Pharmacopeia (USP) for monitoring the ethylene oxide process is Bacillus atrophaeus.
Calibration
Comparison of a measurement system or device of unknown accuracy to a measurement system or device of known accuracy (traceable to national standards) to detect, correlate, report, or eliminate by adjustment, any variation from the required performance limits of the unverified measurement system or device. All critical measuring devices utilized on ethylene oxide sterilizers which may impact the quality of the process are calibrated to traceable national or international standards.
Chamber
Enclosed area which only accommodates sufficient product to fill the sterilizer. EO chambers are constructed of steel or stainless steel and are designed to withstand the extreme pressures and elevated temperatures utilized in the EO sterilization process.
Commissioning
Obtaining and documenting evidence that equipment has been provided and installed in accordance with its specification and that it functions within predetermined limits when operated in accordance with operational instructions. All EO sterilization equipment which complies with ANSI/AAMI/ISO 11135 is commissioned.
Conditioning
Treatment of product within the sterilization cycle, but prior to sterilant admission, to attain a predetermined temperature and relative humidity. This part of the sterilization process may be carried out either at atmospheric pressure or under vacuum (see also preconditioning).
Critical Parameters
Parameters identified as being essential to the sterilization process and requiring monitoring.
Culture Conditions
Stated combination of conditions, including the growth medium with the period and temperature of incubation, used to promote germination, outgrowth, and/or multiplication of the microorganisms for a biological indicator. This must be specified by the BI manufacturer.
Cycle Completion
Point after completion of the sterilization process at which the sterilization load is ready to be removed from the chamber. In EO sterilization, cycle completion occurs after the final air break phase of the process.
D-Value
Time (expressed in minutes) required to achieve inactivation of 90 percent of a population of a test organism under stated exposure conditions. Also referenced as the D10 Value or decimal reduction value.
Development
Process of refining a prototype design or process to meet established product criteria.
Differential Pressure
Difference in pressure between two adjacent pressure points. Differential pressures are utilized when calculating relative humidity and sterilant concentrations in EO sterilization.
Exposure Time
Time for which the sterilizer chamber is maintained within the specified range for temperature, sterilant concentration, pressure, and relative humidity. Also may be referred to as the time for which a medical device (load) is exposed at the specified sterilizing conditions.
Failure
Event in which a component does not perform one or more of its required functions within the specified limits under specified conditions.
Failure Analysis
Logical, systematic examination of an item to identify and analyze the probability, causes, and consequences of potential and real failures.
False Negative
Test result where a true positive was interpreted as negative, where microbial growth was present but not detected, or where viable microorganisms failed to grow.
False Positive
Test results where a true negative was interpreted as a positive, where growth resulted from extraneous microbial contamination, or where turbidity arose from an interaction between the sample and the test medium.
Final Package
Primary containment system, excluding shelf-cartons and shipping containers, that protects contents to the intended level over a specific period of time (i.e., a barrier to physical, microbial, or chemical challenges).
Flushing
Procedure by which sterilant is removed from the load and chamber by either multiple alternate admissions of filtered air or inert gas and evacuations of the chamber or continuous passage of filtered air or inert gas through the load and chamber.
Gas Filter
Nonshedding porous article placed in gas lines to remove particulate matter from gas streams prior to injection into the sterilizer.
Growth Promotion Test
Test performed to demonstrate that media will support microbial growth. For all sterility test performed in support of EO sterilization, all media are growth promotion tested per the requirements and procedures listed in USP.
Health Care Product
Term encompassing medical devices, medicinal products (pharmaceuticals and biologics), and in vitro diagnostics.
Inactivation
Loss of the ability of microorganisms to grow and/or multiply under specified culture conditions.
Indicator
Combination of the indicator agent and its substrate in the form in which it is intended to be used.
Inoculated Carrier
Carrier on which a defined number of test organisms have been deposited.
Installation Qualification
Obtaining and documenting evidence that equipment has been provided and installed in accordance with its specifications and that it functions within predetermined limits when operated in accordance with the operational instructions.
Installation Test
Series of checks and tests performed after installation of a sterilizer in the place of use.
Manufacturer
Natural or legal person packaging or sterilizing a medical device.
Medical Device
Any instrument, apparatus, appliance, material, or other article, whether used alone or in combination, including the software necessary for its proper application intended by the manufacturer to be used for human beings for the purpose of:
- diagnosis, prevention, monitoring, treatment, or alleviation of disease;
- diagnosis, monitoring, treatment, alleviation of, or compensation for an injury or handicap;
- investigation, replacement, or modification of the anatomy or of a physiological process;
- control of conception; and which does not achieve its principal intended action in or on the human body by pharmacological, immunological, or metabolic means, but which may be assisted in its function by such means.
Microbial Barrier
Ability of the packaging system to prevent the ingress of microorganisms under specified conditions.
Microbiological Challenge
Biological indicators, biological-indicator test packs, or inoculated product that contain known populations of microorganisms and can be used in testing sterilization cycles.
Negative Sterility Test
Sterility test samples which do not exhibit detectable microbial growth after incubation.
Nominal Population
The stated number of microorganisms on a biological indicator.
Overkill Sterilization Process
Process which is sufficient to provide at least a 12-logarithmic reduction or 12 D inactivation of an appropriately resistant biological indicator with an established D value.
Package Integrity
Unimpaired physical condition of a final package.
Parametric Release
Declaring product as sterile based on physical and/or chemical process data rather than on the basis of sample testing or biological indicator results.
Performance Qualification
Obtaining and documenting evidence that the equipment, as commissioned, will produce acceptable product when operated according to the processing specifications.
Positive Sterility Test
Sterility test samples which exhibit detectable microbial growth after incubation.
Preconditioning
Treatment of product prior to the sterilization cycle in a room or chamber to attain specified limits for temperature and relative humidity.
Preconditioning Area
Either a chamber or a room in which preconditioning occurs.
Primary Packaging
Element of the packaging system that maintains the sterility of the product.
Process Challenge Device
Object which simulates the worst case of conditions in the items of the goods to be sterilized.
- The device is so constituted that a biological indicator can be arranged in the place most difficult for the sterilant to reach. The design of the process challenge device depends on the kind of goods to be sterilized and the sterilization procedure. The biological indicator should not interfere with the function of the process challenge device.
- In some process challenge devices, an inoculated carrier may be used in place of a biological indicator.
Process Development
Documented program of studies which are performed to define the sterilization process based upon the product/packaging/loading pattern and/or equipment limitations.
Process Lethality
Capability of the sterilization process to destroy microorganisms.
Process Qualification
Obtaining and documenting evidence that the sterilization process will produce acceptable health care products.
Product
Generic term used to describe raw materials, intermediate products, subassemblies, and finished medical devices.
Product Compatibility
Ability of the sterilization process to achieve the intended results without detrimental effect on the product.
Product Qualification
Obtaining and documenting evidence that the health care product will be acceptable for its intended use after exposure to the sterilization process.
Qualification
Documented evidence that all prescribed design and performance requirements are met.
Recommissioning
Repetition of part of all of the commissioning test requirements for the purpose of reconfirming process reliability.
Reference Load
Specified sterilization load made up to represent the most difficult combination of products to be sterilized.
Revalidation
Repetition of part of all of the validation test requirements for the purpose of reconfirming process reliability.
Self-contained Biological Indicators
Biological indicator presented in such a way that the primary pack, intended for incubation, contains the growth medium required for recovery.
Simulated Product Load
Load that is used as an alternative to the actual product load and that represents an equal or greater challenge to the process.
Sterilant
Microbicidal agent in the physical form in which it is active.
Sterilant Injection Stage
Stage beginning with the first introduction of sterilant into the chamber and ending whenever the set operating pressure has been attained.
Sterilant Injection Time
Duration of the sterilant injection stage.
Sterilant Removal Time
Portion of the sterilization cycle in which sterilant is removed from the chamber and sterilization load, but not necessarily desorbed from individual products.
Sterile
Free from viable microorganisms. In practice, no such absolute statement regarding the absence of microorganisms can be proven (see sterilization).
Sterility
State of being free from viable microorganisms. In practice, no such absolute statement regarding the absence of microorganisms can be proven (see sterilization).
Sterility Assurance Level (SAL)
Probability of a viable microorganism being present on a product unit after sterilization. SAL is normally expressed at 10-n.
Sterility Test
Test performed to determine if viable microorganisms are present.
Sterilization
Validated process used to render a product free from viable microorganisms. In a sterilization process, the nature of microbial death is described by an exponential function. Therefore, the presence of viable microorganisms on any individual item can be expressed in terms of probability. While this probability may be reduced to a very low number, it can never be reduced to zero. The probability can be expressed as a sterility assurance level (SAL).
Sterilization Compatibility
Ability of the packaging material to both withstand the intended sterilization process and to allow attainment of the required conditions for sterilization within the final pack.
Sterilization Cycle
Defined sequence of operational steps designed to achieve sterilization that is carried out in a sealed chamber. Specifically for EO sterilization, the treatment in a sealed chamber comprising air removal, conditioning, injection of sterilant, exposure to ethylene oxide, and removal of ethylene oxide.
Sterilization Load
Goods that are to be or have been sterilized simultaneously in the same sterilization chamber.
Sterilization Process
All treatments which are required to accomplish sterilization, including preconditioning, the sterilization cycle, and aeration.
Sterilization Process Development
Studies conducted to develop a reproducible process by which the product can be sterilized to the desired probability of a nonsterile unit (PNSU) without damage.
Terminal Sterilization
A process whereby a product is sterilized in its final container and which permits the measurement and evaluation of quantifiable microbial lethality.
Test of Sterility
Test performed to determine the fraction of product units (or portions thereof) which are scored positive when subjected to defined culture conditions.
Test Organism
Microorganisms for the manufacture of inoculated carriers.
Usable Sterilizer Chamber Volume
Space inside the sterilizer chamber which is not restricted by fixed or mobile parts (loading units, pallets, etc.) and which is consequently available to accept the sterilization load. This is expressed in terms of height, width, and depth.
User
Person making use of the medical device.
Validation
A documented procedure for obtaining, recording, and interpreting the results needed to show that a process will consistently yield a product complying with predetermined specifications. Validation is considered as a total process that includes written protocol, evidence that the equipment as installed meets design criteria and specifications (equipment qualification), use of calibrated instruments to collect data, and evidence that the equipment can deliver the process within specified tolerances under established operating conditions and is reproducible as demonstrated by replicate runs and process challenges (performance qualification).