Dosimeter Comparison Chart
Contact us to find out more about our global contract sterilization, testing, and validation services.
Contact UsDosimeter Comparison Chart
Attribute | Alanine Tape Tab Dosimeter | Red Perspex 4032 Dosimeter |
---|---|---|
Dosimeter Description | Alanine pellets containing alanine. For Tape Tab: Pellet is placed into a film package with a bar code, allowing full traceability of the dosimeter. | Polymethylmethacrylate (PMMA) moulded sheets cut to size and hermetically sealed in a film package containing a barcode. Dosimeter must be removed from it’s packing to use.; |
Dosimeter Dimensions | Alanine Pellet Dimensions
|
|
Dosimeter Response | For Loose Pellets: Use a holder to insert the alanine dosimeter into a spectrometer. For Tape Tab: Dosimeter contains crystalline alanine and registers the absorbed dose by the formation of alanine-derived free radicals. | Induction of chemical reactions resulting from irradiation, which create absorption bands in the visible and/or UV regions of the spectrum. Optical absorbance is determined at appropriate wavelengths within these radiation-induced absorption bands and is quantitatively related to absorbed dose. |
Measurement Instrumentation | Identification and measurement of the alanine-derived free radicals are performed by Electron Paramagnetic Resonance (EPR) spectrometer. | Spectrophotometer Micrometer |
Dosimeter Measurement Process | Insert alanine dosimeter into spectrometer, system automatically transfers dosimeter bar code into software and begins taking measurement, dose calculated. | Users remove dosimeter from packet. Place dosimeters into holder. Place holder into spectrophotometer and absorbance transferred into software. Remove dosimeters from holder and read thickness of dosimeters, one at a time. Thickness automatically transferred into software, dose calculated. |
Influence Quantities | Temperature, pellet mass | Temperature, humidity, dose rate, post-irradiation response |
Typical Dosimeter System Uncertainty | ± 4% | ± 6% |
Typical Dose Range | 0.01-100 kGy (0.01kGy to 100kGy may also be achieved) | 5-50 kGy |