- Pallets pass in front of the X-ray beam on two levels
- Pallets can pass in front of the beam multiple times with each pallet quadrant being exposed equally
- Conveyance mechanism moves pallets in front of the beam at a speed of 0.5 to 4.5 m/min. Speed regulation is independent of the number of pallets loaded on this section of the conveyor
- The beam area or treatment area conveyor is 3 m long (2 such sections – top and bottom)
- Conveyor mechanism provides a gap closing conveyor sequence so that minimal gap between pallets is constant
The product is treated from 2 sides and the bottom pallet has to get the same double side treatment on the upper position, forming a 4 passes system. The depth of the irradiation is made on the long side of the pallets (1.2 m).
In order to give dose flexibility, the treatment is given by ~2kGy increments, corresponding to one full sequence of 4 passes in front of the X-ray target.
When multiple dose increments are required, i.e.,2, the corresponding container will travel 48 times (12 x 4) around the central wall in order to alternate the face which is presented to the X-ray. After every 2 passes it is changing the level from bottom to upper or contrary.
Related TechTips
Considerations When Establishing a Maximum Sterilization Dose for Radiation Processing
Per ISO 11137, products sterilized by X-ray, gamma, and E-beam irradiation require a documented and validated dose range. A dose range is composed of: A minimum dose based on the microbiologic considerations (type and quantity of organisms and their re
Radiation Sterilization Master File Pilot Program FAQ
Q1: What is the Radiation Sterilization Master File Pilot Program? A:The Radiation Sterilization Master File Pilot Program (Pilot Program) is a voluntary program that intends to allow companies that terminally sterilize single-use medical devices using
Transferring of the Verification Dose from Electron Beam Technology to Photon Technology (Gamma or X-ray)
Background A verification (dose audit) or sterilization dose experiment requires a narrow and specific dose range to be applied to the samples used for tests of sterility as evidence that the routine minimum dose remains appropriate for delivering the
Activation Assessment (X-ray and electron beam)
Background Radioactive material of natural origin is abundant and varies widely in both type and amount. The energy from these materials, plus that of cosmic or cosmogenic origin, is called background radiation. Artificial radioactivity is the result o